Lab 2: Fluorescence Microscopy

Goals

In this module, I am interested in:

  • Building and optimizing a fluorescent microscope
  • Learning to control the microstepper
  • Learning to talk to and control the camera
  • Learning more about 3d deconvolution
    • Test how it deconvolution is improved with actual PSF vs. theoretical PSF of the microscope

Methodology

Materials

  • Pololu A4983 Stepper Motor Driver Carrier with Voltage Regulators[1]
  • Thorlabs 8 mm Travel Stepper Motor Actuator[2]
  • Matlab
  • Matlab Image Acquisition Toolbox[3]
  • Matlab Allied Vision Technologies Adaptor[4]
  • National Instruments USB Data Acquisition[5]
  • Epi-fluorescence Microscope[6] with parts from Thorlabs
  • Submicron beads, small enough to be considered point sources

Deconvolution Methods for 3D Fluorescence Microscopy Images[7]

Microscope construction

  • The microscope was a reanimation of the teaching microscope from the 20.309 lab. We changed the lenses a bit.
  • The scope contains a 8.33x beam expander and 40x objective as of now. The objectives are changeable, but require some adjustment of the 200mm tube lens.

Laser Alignment

  • The neutral density filter ensures we do not burn our eyeballs while aligning the laser. 532 nm protective goggles should be worn when appropriate.
  • T-shirt targets were used as guides to center the expanded beam with the mirrors.
  • With barrier filter still in place, we used a fluorescent reference slide and Steve's Super Special Software (or something like that) called Capture Image or something to look at the intensity histogram and redirect the mirror just prior to the CCD camera to maximize light homogeneity.
  • Profit.

PSF Scanning

Results

Code

References

  1. A4983 Stepper Motor Driver Carrier with Voltage Regulators
  2. 8 mm Travel Stepper Motor Actuator
  3. Mathworks Image Acquisition Toolbox
  4. Matlab Allied Vision Technologies Adaptor
  5. National Instruments USB Data Acquisition 6210
  6. Lab_Manual:Optical_Microscopy
  7. Deconvolution Methods for 3D Fluorescence Microscopy Images - WUSTL 2006