Difference between revisions of "Assignment 2 Part 3: Build an epi-illuminator for your microscope"

From Course Wiki
Jump to: navigation, search
(Fluorescence imaging)
(Flat field correction)
 
(23 intermediate revisions by 3 users not shown)
Line 7: Line 7:
 
[[Image:20.309 130911 YourMicroscope.png|thumb|260px|right|Fluorescence microscope block diagram.]]
 
[[Image:20.309 130911 YourMicroscope.png|thumb|260px|right|Fluorescence microscope block diagram.]]
  
In this part of the lab, you will add epifluorescence imaging capability to your microscope, make images of fluorescent samples, and process the images to correct artifacts caused by nonuniform illumination. This week, you will test out your new fluorescence microscope by imaging fluorescent microspheres of several sizes. Next week you'll image mammalian cells stained with a fluorescent dye. To make the correction for nonuniform illumination, you will also make images of a uniform fluorescence reference slide and a dark image with the illuminator turned off.
+
It's time to start building! In this part of the assignment, you're going to modify your microscope to add the fluorescence capabilities we talked about in Part 2. You'll then test out these new capabilities by recording images of fluorescent microspheres. Finally, you'll process those images using a flat-field correction to adjust for non-uniform illumination. Once you've completed this assignment, you should be ready for next week when you'll make beautiful images of stained actin filaments inside of mammalian cells!
 +
 
 +
<br/>
 +
<br/>
 
<br/>
 
<br/>
  
==Laser safety==
+
==High power light source safety==
 
{{Template:Safety Warning|message=
 
{{Template:Safety Warning|message=
In this part of the lab, you will use a 5mW, 532 nm laser with focusing optics. You must attend the safety lecture before you work with the laser. See an instructor if you missed the lecture. Do not begin working with the laser until you thoroughly understand how to use it safely.
+
In this part of the lab, you will use a ~3W green LED with focusing optics. While LEDs are more diffuse than lasers, they still have the potential to cause harm to your eyes. You must attend the safety lecture before you work with high power LEDs in lab. See an instructor if you missed the lecture. Do not begin working with the green LED until you thoroughly understand how to use it safely.
  
Stop working and ask an instructor immediately if you have any questions about working with lasers safely.  
+
Stop working and ask an instructor immediately if you have any questions about working with lasers or high-power LEDs safely.  
  
When you work with lasers, keep these laser safety best practices in mind:
+
When you work with high power light sources, keep these best practices in mind:
  
* Remove reflective clothing items and jewelry when working with lasers.
+
* Remove reflective clothing items and jewelry.
* Use the lowest possible power during alignment or adjustment.
+
* Use the lowest possible power during alignment or adjustment - for LEDs, that means work at a low current level, for lasers, that means using a neutral density filter to reduce the laser power.
* Do not use reflective tools when the laser is on.
+
* Do not use reflective tools when the light source is on.
 
* Know the beam path at all times.
 
* Know the beam path at all times.
 
** Use a stop to prevent uncontained beams.
 
** Use a stop to prevent uncontained beams.
 
** Secure the stop so it will not move accidentally.
 
** Secure the stop so it will not move accidentally.
 
** Keep your eyes out of the plane of the beam, which his normally a plane just above the optical table.
 
** Keep your eyes out of the plane of the beam, which his normally a plane just above the optical table.
* Disable lasers when they are not in use.
+
* Disable light sources when they are not in use.
 
* Use the correct safety goggles.
 
* Use the correct safety goggles.
** Always check the marking on the goggles to ensure they are appropriate for the laser you are working with.
+
** Always check the marking on the goggles to ensure they are appropriate for the light sources you are working with.
** Laser safety goggles are a last line of defense, not an excuse for unsafe practices.
+
** Safety goggles are a last line of defense, not an excuse for unsafe practices.
* Do not operate a laser unless the flashing laser warning sign is on.
+
* Do not operate a high-power light source unless the flashing laser warning sign is on.
  
 
Some specific ways to work safely with the 20.309 fluorescence microscope:
 
Some specific ways to work safely with the 20.309 fluorescence microscope:
  
* Use an OD = 2 neutral density (ND) filter to attenuate the laser near the source.
+
* Use an OD = 2 neutral density (ND) filter to attenuate a laser near the source.
 
** "OD" is an abbreviation for "optical density," which is log<sub>10</sub> of the power transmitted by the filter divided by the incident power.  
 
** "OD" is an abbreviation for "optical density," which is log<sub>10</sub> of the power transmitted by the filter divided by the incident power.  
 
** "ND" means that the OD is equal for all wavelengths.
 
** "ND" means that the OD is equal for all wavelengths.
** The ND filter attenuates the power of the laser by a factor of 10<sup>2</sup>, to about 50 &mu;W, which is completely safe.
+
** The ND filter attenuates the power of the light by a factor of 10<sup>2</sup>.
 
** Mount the ND filter securely so there is no chance that it will get knocked out of place accidentally.
 
** Mount the ND filter securely so there is no chance that it will get knocked out of place accidentally.
* The inverted microscope design directs laser light upward, toward the ceiling. Do not lean over the objective mount. Use a stop to prevent the beam from propagating above your microscope.
+
* The inverted microscope design directs illumination light upward, toward the ceiling. Cover the objective and sample with a box and avoid looking directly at the bare objective when the light source is at full power. Do not lean over the objective mount. Use a stop to prevent the beam from propagating above your microscope.
* Confine beams inside lens tubes.
+
* Confine beams inside lens tubes or plastic pipes.
* Disable the laser when it is unattended or stored by removing a battery, disconnecting the power source, or removing the key.
+
* Disable the light source when it is unattended or stored by removing a battery, disconnecting the power source, or removing the key.
 
* Safety goggles for the laser in this lab should have an OD of at least 2 at 532 nm.
 
* Safety goggles for the laser in this lab should have an OD of at least 2 at 532 nm.
 
** There are (at least) two different kinds of safety goggles in the lab that are appropriate for this lab. One has an OD of 2 at 532 nm and the other has an OD of 5. You can see the beam with the OD = 2 goggles, but the power is reduced to a safe level. This makes the OD 2 laser goggles very useful for making adjustments that must be done at full power.
 
** There are (at least) two different kinds of safety goggles in the lab that are appropriate for this lab. One has an OD of 2 at 532 nm and the other has an OD of 5. You can see the beam with the OD = 2 goggles, but the power is reduced to a safe level. This makes the OD 2 laser goggles very useful for making adjustments that must be done at full power.
 
}}
 
}}
  
==Add laser illuminator, dichroic mirror, and barrier filter==
+
==Add fluorescence capabilities to your microscope==
[[Image:20.309 130816 FluoMicroscope Annotated.png|thumb|right|400px|Layout of fluorescence microscope.]]
+
 
# Go over your design with one of the instructors before you start building.
+
===Add illuminator and dichroic mirror===
# Use cage rods to construct a cage of appropriate size for the beam expander and excitation tube lens (L3, L4, and L5).
+
[[Image:LedEpifluorescenceMicroscope.png|thumb|center|400px|Layout of fluorescence microscope.]]
#* For maximum flexibility in positioning, removing, and reinstalling optics, use 3 cage rods instead of 4. Use a cage plate (CP02) for each lens. Mount the lens in a 1/2" lens tube (SM1L05) so it can be easily removed or installed.
+
Use your design from Part 2 and the example microscope to guide you in building the fluorescence part of your microscope. Here are some useful tips to keep in mind:
# Add the two turning mirrors to the end of the cage.
+
# Go over your illuminator design with one of the instructors before you start building if there are any parts you're unsure about.
#* Use a thick cage plate (CP02T) only to join 2 cages. (You will use a total of two CP02T plates in your microscope.) Don't use CP02T to mount lenses &mdash; there are not enough of them in the lab to use this way. It is easy to tell the difference. CP02T has 8 set screws and is roughly twice as thick as the CP02; CP02 has 4 set screws.
+
# For maximum flexibility in positioning, removing, and reinstalling optics, use 3 cage rods instead of 4.  
# Use a cube optic mount (B5C) to mount dichroic mirror DM on a kinematic platform (B4C) . Place the kinematic platform in the cage cube (C6W).  
+
# You'll want the 75 mm excitation tube lens (L3) to be as close as possible to the objective. Use two retaining rings to secure the lens at the inlet of the C6W cage cube.  
#* The first surface of the dichroic should face the laser.
+
#*[[Image:ExcitationTubeLensC6W.jpg|center|thumb|200px|Mount the excitation tube lens in the C6W cage cube.]]
 +
# In the second half of the semester, you will upgrade your microscope (even more!) to make two-color fluorescence images. To avoid having to disassemble your microscope, you'll leave an extra cage cube (C6W) in between your green LED and the excitation tube lens. Mounting the cube to a post will also provide some stability and rigidity to your illuminator.
 +
#* As before, the B1C cage cube cover plate has a shallow 8-32 threaded hole. Be sure to use 2 washers in between your TR2 post and the B1C base to be able to tighten the screw adequately without it getting stuck.
 +
#* [[Image:FilterCubePostAssembly.png|center|thumb|200px|Mounting a second filter cube to a TR2 post and securing it in a PH2 post holder will provide the illuminator stability and allow you to expand your microscope for two colors later in the class.]]
 +
#* Use the tiny 1/4" cage rods (ER025) to connect the two C6W cages together.  
 +
#*[[Image:FilterCubeConnection.png|center|thumb|200px|Use ER025 cage rods to connect two cage cubes.]]
 +
# Mount the green excitation filter (EX1: Chroma Technologies ET550/20) in a 1/2" lens tube (SM1L05). Use two retaining rings like you did in Assignment 1 to mount your 25 mm lens for the trans illuminator.
 +
# Mount the aspheric lens (L4) in a 1/2" lens tube (SM1L05). The 20 mm asphere (L4) looks very similar to the 25 mm plano convex lenses used previously, so make sure to label it appropriately.
 +
# Use some plastic pipe as a sleeve around the bright green LED to protect your eyes.
 +
#*[[Image:PlasticPipeSleeve.jpg|center|thumb|200px|Use plastic pipe to protect your eyes from the bright LEDs.]]
 +
# The dichroic mirror should already be mounted to a (B4C) platform using an FFM1 clamp. Insert the B4C platform into the cage cube (C6W). Be careful not to get fingerprints on the mirror.
 +
#* The first surface of the dichroic should face the excitation source.
 
#** Some dichroics have an arrow indicating the first surface. If not, to ascertain which surface has the coating, watch the reflection of a corner of a piece of lens paper as it touches the mirror. On the first surface, the corner and its reflection will appear to touch. Held the other way, the corner and its image will appear a few millimeters apart.
 
#** Some dichroics have an arrow indicating the first surface. If not, to ascertain which surface has the coating, watch the reflection of a corner of a piece of lens paper as it touches the mirror. On the first surface, the corner and its reflection will appear to touch. Held the other way, the corner and its image will appear a few millimeters apart.
#* Mount the dichroic so that the first surface lies on exactly a diameter of the B5C mount.
+
#* [[Image:FFM1.png|center|thumb|200px|Dichroic mirror mount]]
#* The mounting bracket should not stand in the way of the rotation of the kinematic stage.
+
#* Use the clear plastic #4/40 screws affix the dichroic mirror without deforming nor scratching its surface. 
+
#* Hold the screwdriver at the tip to avoid slipping and scratching the dichroic. The screwdriver will instantly scratch the dichroic.
+
#* [[Image:20.309 130816 DichroicMirrorMount2.png|center|thumb|600px|Dichroic mirror mount]]
+
 
#* Ask an instructor for help if you need to clean a dichroic mirror or barrier filter. Dichroics have delicate, exposed coatings and must be cleaned with extra care.
 
#* Ask an instructor for help if you need to clean a dichroic mirror or barrier filter. Dichroics have delicate, exposed coatings and must be cleaned with extra care.
# Use four black, plastic #4/40 screws to hold the cube optic mount (B5C) on the cage cube (C6W).
+
# Use four black, plastic #4/40 screws to hold the dichroic mount (B4C) on the cage cube (C6W).
 
#* Tighten the screws enough so that the cube optic mount holds its adjustment, but can still be rotated.
 
#* Tighten the screws enough so that the cube optic mount holds its adjustment, but can still be rotated.
#* Make sure to block the (small percentage of) excitation laser light that will be transmitted through the dichroic mirror with some black lab tape affixed to the B5C cube optic mount.
+
#* Cover the back open port of the C6W cage cube to prevent stray light from getting to your camera.
 +
# Insert the emission filter into the emission path of your microscope (between the mirror and the tube lens). In the later part of the semester, we will add a second color LED to our microscope. The particular emission filter we'll use will work with either green or blue excitation.
  
===Align the laser illumination path===
+
===Align the illumination path===
  
# Remove L3, L4, L5, and the objective lens L1.
+
# Use the lowest possible LED current when aligning. Make sure to use some plastic pipe as a sleeve around the LED to reduce stray light, and to cover the back of the C6W cube with electrical tape. Don't hesitate to ask an instructor for help during the alignment process.
# Mount the laser.
+
# Adjust the dichroic mirror DM so that the beam will enter the middle of the objective lens.  
# Insert a neutral density filter (ND filter) between the laser and M2 to reduce the laser power to a safe level for adjustment.
+
#* Remove your objective and cover the opening with an opaque screen (i.e. a kimwipe or lens paper).
#* Remove jewelry. Turn of the laser before using reflective tools.
+
#* Connect your LED to the power supply. Adjust the current limit to zero, and the voltage to about 6V, and then enable the power. Gently turn up the LED power until the green light turns on. This should be less than 0.1A. (Never exceed 1A of current through these LEDs.)
# Turn on the laser. Use a beam stop until the laser position is set. Adjust the laser position so that the laser shines near the middle of M2.
+
#* Rotate the B4C platform holding the dichroic mirror until you see some light hitting your screen. Continue rotating until the pattern is centered. You may also need to use the adjustment screws on the B4C platform to center the beam in both X and Y.
#* You can use a DG10 mounted frosted-glass alignment disk with a pinhole to best center the laser beam onto M2.
+
#* Replace the objective.
# Adjust M2 and M3 to center the laser light in the cage. 
+
# Use a fluorescence reference slide maximize the illumination intensity and center the field of view.  
#* Use two CPA1 alignment targets to gauge beam alignment.
+
#* Start up the <tt>UsefulImageAcquisition</tt> tool and set the exposure to an appropriate level.  
#* [[Image:20.309 130816 AlignPinholes 4.png|center|thumb|600px|Using pinholes and irises to optimize initial microscope alignment. Notice that this person is breaking one of the laser safety rules. Which one?]]
+
#* Bring the reference slide approximately into focus.
# Adjust the dichroic mirror DM so that the beam will enter the middle of the objective lens.
+
#* Use the bottom mirror (not the dichroic) to center the illumination on the field of view of the camera. ''Tip:'' Check the "Stretch Contrast" box on the software to highlight the bright and dark regions of the image.  
#* Use a DG10 mounted frosted-glass alignment disk with a pinhole to show the optical center of the vertical beam path.
+
#* Using the image histogram as your guide, rotate the dichroic mirror until the intensity is maximized. If necessary, you may also need to adjust the tilt of the mirror using the adjustment screws on the B4C base.
#* Verify that the centered beam is perpendicular to the floor. If the beam is at an angle, verify that the dichroic mirror DM is mounted on a diameter of the rotating mount.
+
#* Next align the position of the LED relative to its collimating lens. Loosen the set screws holding the LED in place, and slide the LED assembly forward and backward along the cage rods until the intensity is maximized (again, use the histogram to gauge the intensity values).  
# Insert the beam expander lenses, L3 and L4. Adjust the separation between L3 and L4 to achieve a collimated beam.
+
#* Once you've found the optimal LED position, re-tighten it's set screws. You may need to re-check the dichroic mirror alignment now that you have altered the LED position.  
#* It may be necessary to make small adjustments to M2 and M3 to recenter the beam.
+
#* Finally, if necessary, readjust the bottom mirror's alignment to center the illumination on the field of view of the camera (again, stretching the contrast is helpful here).
# Replace the excitation tube lens L5 and the objective lens L1. Adjust the position of L5 for best beam collimation (you can try to focus it far away).
+
#* The product of beam divergence and diameter is constant. L5 and L1 shrink the beam, causing increased divergence. The beam emerging from the objective will likely appear to grow in size as it propagates, even when the lenses are in their optimal positions.
+
#* It may be necessary to make small adjustments to M2 and M3 to recenter the beam.
+
# Use a fluorescence reference slide to center the field of view and to optimize the uniformity of illumination.  
+
#* Make sure to use an ND filter when you use the fluorescence reference slide. The slide bleaches quickly at high intensities and you will not get a good image.
+
#* Put a barrier filter in the afocal part of the imaging path.
+
#* Center the camera's FOV in the objective's FOV.
+
#* If the laser light emerges from the objective at an angle, double-check your alignment and dichroic mounting.
+
#* The barrier filter has delicate, exposed metal coatings. Ask an instructor for help if you need to clean it.
+
  
 
==Fluorescence imaging==
 
==Fluorescence imaging==
Line 97: Line 99:
  
 
* Read [[Flat-field correction|this page]] to learn about the flat-field correction.
 
* Read [[Flat-field correction|this page]] to learn about the flat-field correction.
* Reference Images: Take the reference image as close as possible in time to the sample images. Don't make any adjustments to your microscope between capturing the reference image and the sample image. For example, every time you move a mirror or re-align the laser, you will change the illumination profile and you must take a new reference. Adjusting the camera exposure and gain between recording the reference and sample images is okay.  
+
* Reference images: Take the reference image as close as possible in time to the sample images. Don't make any adjustments to your microscope between capturing the reference image and the sample image. For example, every time you move a mirror or re-align the laser, you will change the illumination profile and you must take a new reference. Adjusting the camera exposure and gain between recording the reference and sample images is okay.  
* Dark images: Each time you record an image (reference or sample), make sure to take a corresponding dark image using the EXACT SAME camera settings (''i.e. use the exact same Exposure Time and Gain settings you had chosen for your reference/sample image''). This is the only valid way to subtract the correct dark value from your reference/sample image. Use 12-bit imaging mode to get the best results.  
+
* Dark images: Each time you record an image (reference or sample), make sure to take a corresponding dark image using the EXACT SAME camera settings (''i.e.'' use the exact same <tt>Exposure Time</tt> and <tt>Gain</tt> settings you had chosen for your reference/sample image). This is the only valid way to subtract the correct dark value from your reference/sample image. Use 12-bit imaging mode to get the best results.  
 
* Read through [[Recording, displaying and saving images in MATLAB|this page]] to refresh you memory on saving in MATLAB. We recommend using the <tt>save</tt> command to save data as a .mat file that you can load back into MATLAB later. This makes it easy to reanalyze your images if necessary. You could also save in an image file format with <tt>imwrite</tt>. Convert the image to 16-bit, unsigned integer format with the correct range before saving.  
 
* Read through [[Recording, displaying and saving images in MATLAB|this page]] to refresh you memory on saving in MATLAB. We recommend using the <tt>save</tt> command to save data as a .mat file that you can load back into MATLAB later. This makes it easy to reanalyze your images if necessary. You could also save in an image file format with <tt>imwrite</tt>. Convert the image to 16-bit, unsigned integer format with the correct range before saving.  
  
 
===Take images===
 
===Take images===
 
# Record a reference image  
 
# Record a reference image  
#* Put in an ND filter and take an image of the reference slide with the 40X objective.  
+
#* Using a low LED intensity, take an image of the reference slide with the 40X objective.  
#* Turn off the laser and record a dark image (without changing any camera settings!)
+
#* Turn off the LED illumination and record a dark image (without changing any camera settings!)
#* Use MATLAB to generate a histogram of to be certain that the image is exposed correctly.
+
#* Use MATLAB to generate a histogram of each image to be certain that the image is exposed correctly.
# Remove the ND filter and image the samples:
+
# Record fluorescence images of the following two samples:
 
#* 0.84 &mu;m beads and a corresponding dark image
 
#* 0.84 &mu;m beads and a corresponding dark image
 
#* 3.26 &mu;m beads and a corresponding dark image
 
#* 3.26 &mu;m beads and a corresponding dark image
 
+
#* Keep in mind:
* For each of these samples remember to make sure the image is exposed correctly, and to take a corresponding dark image
+
#** You may need to increase your LED intensity by increasing the current limit on the power supply. '''Never allow the current go above 1A!'''
* Adjust the gain (to zero) and exposure of the camera to get the best picture.
+
#** For each sample, remember to make sure the image is exposed correctly, and to take a corresponding dark image.
 +
#** Adjust the gain (to zero) and increase the exposure of the camera to get the best picture.
  
 
==Flat field correction==
 
==Flat field correction==
Line 118: Line 121:
 
#* ''Tip'': If all your images are appearing dark in MATLAB you probably have not rescaled your image properly. Read through the ''displaying images'' section of [[Recording, displaying and saving images in MATLAB|this page]] if you need a refresher.
 
#* ''Tip'': If all your images are appearing dark in MATLAB you probably have not rescaled your image properly. Read through the ''displaying images'' section of [[Recording, displaying and saving images in MATLAB|this page]] if you need a refresher.
  
{{Template:Assignment Turn In|message = <br/>
+
{{Template:Assignment Turn In|message =  
 
# For each bead sample, include the original, reference, and flat-field corrected images in your lab report. In the caption note the exposure and gain settings used for each image.
 
# For each bead sample, include the original, reference, and flat-field corrected images in your lab report. In the caption note the exposure and gain settings used for each image.
 
# For one set of images (either the 0.84 or 3.6 &mu;m beads and their corresponding dark and reference images), include the MATLAB code you used to calculate the flat-field correction.
 
# For one set of images (either the 0.84 or 3.6 &mu;m beads and their corresponding dark and reference images), include the MATLAB code you used to calculate the flat-field correction.

Latest revision as of 19:21, 21 February 2020

20.309: Biological Instrumentation and Measurement

ImageBar 774.jpg


Overview

Fluorescence microscope block diagram.

It's time to start building! In this part of the assignment, you're going to modify your microscope to add the fluorescence capabilities we talked about in Part 2. You'll then test out these new capabilities by recording images of fluorescent microspheres. Finally, you'll process those images using a flat-field correction to adjust for non-uniform illumination. Once you've completed this assignment, you should be ready for next week when you'll make beautiful images of stained actin filaments inside of mammalian cells!




High power light source safety


Warning.jpg In this part of the lab, you will use a ~3W green LED with focusing optics. While LEDs are more diffuse than lasers, they still have the potential to cause harm to your eyes. You must attend the safety lecture before you work with high power LEDs in lab. See an instructor if you missed the lecture. Do not begin working with the green LED until you thoroughly understand how to use it safely.

Stop working and ask an instructor immediately if you have any questions about working with lasers or high-power LEDs safely.

When you work with high power light sources, keep these best practices in mind:

  • Remove reflective clothing items and jewelry.
  • Use the lowest possible power during alignment or adjustment - for LEDs, that means work at a low current level, for lasers, that means using a neutral density filter to reduce the laser power.
  • Do not use reflective tools when the light source is on.
  • Know the beam path at all times.
    • Use a stop to prevent uncontained beams.
    • Secure the stop so it will not move accidentally.
    • Keep your eyes out of the plane of the beam, which his normally a plane just above the optical table.
  • Disable light sources when they are not in use.
  • Use the correct safety goggles.
    • Always check the marking on the goggles to ensure they are appropriate for the light sources you are working with.
    • Safety goggles are a last line of defense, not an excuse for unsafe practices.
  • Do not operate a high-power light source unless the flashing laser warning sign is on.

Some specific ways to work safely with the 20.309 fluorescence microscope:

  • Use an OD = 2 neutral density (ND) filter to attenuate a laser near the source.
    • "OD" is an abbreviation for "optical density," which is log10 of the power transmitted by the filter divided by the incident power.
    • "ND" means that the OD is equal for all wavelengths.
    • The ND filter attenuates the power of the light by a factor of 102.
    • Mount the ND filter securely so there is no chance that it will get knocked out of place accidentally.
  • The inverted microscope design directs illumination light upward, toward the ceiling. Cover the objective and sample with a box and avoid looking directly at the bare objective when the light source is at full power. Do not lean over the objective mount. Use a stop to prevent the beam from propagating above your microscope.
  • Confine beams inside lens tubes or plastic pipes.
  • Disable the light source when it is unattended or stored by removing a battery, disconnecting the power source, or removing the key.
  • Safety goggles for the laser in this lab should have an OD of at least 2 at 532 nm.
    • There are (at least) two different kinds of safety goggles in the lab that are appropriate for this lab. One has an OD of 2 at 532 nm and the other has an OD of 5. You can see the beam with the OD = 2 goggles, but the power is reduced to a safe level. This makes the OD 2 laser goggles very useful for making adjustments that must be done at full power.


Add fluorescence capabilities to your microscope

Add illuminator and dichroic mirror

Layout of fluorescence microscope.

Use your design from Part 2 and the example microscope to guide you in building the fluorescence part of your microscope. Here are some useful tips to keep in mind:

  1. Go over your illuminator design with one of the instructors before you start building if there are any parts you're unsure about.
  2. For maximum flexibility in positioning, removing, and reinstalling optics, use 3 cage rods instead of 4.
  3. You'll want the 75 mm excitation tube lens (L3) to be as close as possible to the objective. Use two retaining rings to secure the lens at the inlet of the C6W cage cube.
    • Mount the excitation tube lens in the C6W cage cube.
  4. In the second half of the semester, you will upgrade your microscope (even more!) to make two-color fluorescence images. To avoid having to disassemble your microscope, you'll leave an extra cage cube (C6W) in between your green LED and the excitation tube lens. Mounting the cube to a post will also provide some stability and rigidity to your illuminator.
    • As before, the B1C cage cube cover plate has a shallow 8-32 threaded hole. Be sure to use 2 washers in between your TR2 post and the B1C base to be able to tighten the screw adequately without it getting stuck.
    • Mounting a second filter cube to a TR2 post and securing it in a PH2 post holder will provide the illuminator stability and allow you to expand your microscope for two colors later in the class.
    • Use the tiny 1/4" cage rods (ER025) to connect the two C6W cages together.
    • Use ER025 cage rods to connect two cage cubes.
  5. Mount the green excitation filter (EX1: Chroma Technologies ET550/20) in a 1/2" lens tube (SM1L05). Use two retaining rings like you did in Assignment 1 to mount your 25 mm lens for the trans illuminator.
  6. Mount the aspheric lens (L4) in a 1/2" lens tube (SM1L05). The 20 mm asphere (L4) looks very similar to the 25 mm plano convex lenses used previously, so make sure to label it appropriately.
  7. Use some plastic pipe as a sleeve around the bright green LED to protect your eyes.
    • Use plastic pipe to protect your eyes from the bright LEDs.
  8. The dichroic mirror should already be mounted to a (B4C) platform using an FFM1 clamp. Insert the B4C platform into the cage cube (C6W). Be careful not to get fingerprints on the mirror.
    • The first surface of the dichroic should face the excitation source.
      • Some dichroics have an arrow indicating the first surface. If not, to ascertain which surface has the coating, watch the reflection of a corner of a piece of lens paper as it touches the mirror. On the first surface, the corner and its reflection will appear to touch. Held the other way, the corner and its image will appear a few millimeters apart.
    • Dichroic mirror mount
    • Ask an instructor for help if you need to clean a dichroic mirror or barrier filter. Dichroics have delicate, exposed coatings and must be cleaned with extra care.
  9. Use four black, plastic #4/40 screws to hold the dichroic mount (B4C) on the cage cube (C6W).
    • Tighten the screws enough so that the cube optic mount holds its adjustment, but can still be rotated.
    • Cover the back open port of the C6W cage cube to prevent stray light from getting to your camera.
  10. Insert the emission filter into the emission path of your microscope (between the mirror and the tube lens). In the later part of the semester, we will add a second color LED to our microscope. The particular emission filter we'll use will work with either green or blue excitation.

Align the illumination path

  1. Use the lowest possible LED current when aligning. Make sure to use some plastic pipe as a sleeve around the LED to reduce stray light, and to cover the back of the C6W cube with electrical tape. Don't hesitate to ask an instructor for help during the alignment process.
  2. Adjust the dichroic mirror DM so that the beam will enter the middle of the objective lens.
    • Remove your objective and cover the opening with an opaque screen (i.e. a kimwipe or lens paper).
    • Connect your LED to the power supply. Adjust the current limit to zero, and the voltage to about 6V, and then enable the power. Gently turn up the LED power until the green light turns on. This should be less than 0.1A. (Never exceed 1A of current through these LEDs.)
    • Rotate the B4C platform holding the dichroic mirror until you see some light hitting your screen. Continue rotating until the pattern is centered. You may also need to use the adjustment screws on the B4C platform to center the beam in both X and Y.
    • Replace the objective.
  3. Use a fluorescence reference slide maximize the illumination intensity and center the field of view.
    • Start up the UsefulImageAcquisition tool and set the exposure to an appropriate level.
    • Bring the reference slide approximately into focus.
    • Use the bottom mirror (not the dichroic) to center the illumination on the field of view of the camera. Tip: Check the "Stretch Contrast" box on the software to highlight the bright and dark regions of the image.
    • Using the image histogram as your guide, rotate the dichroic mirror until the intensity is maximized. If necessary, you may also need to adjust the tilt of the mirror using the adjustment screws on the B4C base.
    • Next align the position of the LED relative to its collimating lens. Loosen the set screws holding the LED in place, and slide the LED assembly forward and backward along the cage rods until the intensity is maximized (again, use the histogram to gauge the intensity values).
    • Once you've found the optimal LED position, re-tighten it's set screws. You may need to re-check the dichroic mirror alignment now that you have altered the LED position.
    • Finally, if necessary, readjust the bottom mirror's alignment to center the illumination on the field of view of the camera (again, stretching the contrast is helpful here).

Fluorescence imaging

Example images and histogram. Top: fluorescent beads; center: reference slide, bottom: image histogram.

In this part of the lab, you will make images of fluorescent microspheres and correct them for nonuniform illumination. In order to do the correction, you will need a reference image and dark image in addition to the image of the sample.

  • Read this page to learn about the flat-field correction.
  • Reference images: Take the reference image as close as possible in time to the sample images. Don't make any adjustments to your microscope between capturing the reference image and the sample image. For example, every time you move a mirror or re-align the laser, you will change the illumination profile and you must take a new reference. Adjusting the camera exposure and gain between recording the reference and sample images is okay.
  • Dark images: Each time you record an image (reference or sample), make sure to take a corresponding dark image using the EXACT SAME camera settings (i.e. use the exact same Exposure Time and Gain settings you had chosen for your reference/sample image). This is the only valid way to subtract the correct dark value from your reference/sample image. Use 12-bit imaging mode to get the best results.
  • Read through this page to refresh you memory on saving in MATLAB. We recommend using the save command to save data as a .mat file that you can load back into MATLAB later. This makes it easy to reanalyze your images if necessary. You could also save in an image file format with imwrite. Convert the image to 16-bit, unsigned integer format with the correct range before saving.

Take images

  1. Record a reference image
    • Using a low LED intensity, take an image of the reference slide with the 40X objective.
    • Turn off the LED illumination and record a dark image (without changing any camera settings!)
    • Use MATLAB to generate a histogram of each image to be certain that the image is exposed correctly.
  2. Record fluorescence images of the following two samples:
    • 0.84 μm beads and a corresponding dark image
    • 3.26 μm beads and a corresponding dark image
    • Keep in mind:
      • You may need to increase your LED intensity by increasing the current limit on the power supply. Never allow the current go above 1A!
      • For each sample, remember to make sure the image is exposed correctly, and to take a corresponding dark image.
      • Adjust the gain (to zero) and increase the exposure of the camera to get the best picture.

Flat field correction

  1. Perform flat-field correction on the images.
    • Divide the image by a normalized version of your reference image minus the dark image (see this page for more detail).
    • Tip: If all your images are appearing dark in MATLAB you probably have not rescaled your image properly. Read through the displaying images section of this page if you need a refresher.


Pencil.png
  1. For each bead sample, include the original, reference, and flat-field corrected images in your lab report. In the caption note the exposure and gain settings used for each image.
  2. For one set of images (either the 0.84 or 3.6 μm beads and their corresponding dark and reference images), include the MATLAB code you used to calculate the flat-field correction.


Navigation

Back to 20.309 Main Page