Assignment 2 Part 3: Build an epi-illuminator for your microscope
Overview
In this part of the lab, you will add epifluorescence imaging capability to your microscope, make images of fluorescent samples, and process the images to correct artifacts caused by nonuniform illumination. This week, you will test out your new fluorescence microscope by imaging fluorescent microspheres of several sizes. Next week you'll image mammalian cells stained with a fluorescent dye. To make the correction for nonuniform illumination, you will also make images of a uniform fluorescence reference slide and a dark image with the illuminator turned off.
Laser safety
Add illuminator, dichroic mirror, and barrier filter
- Go over your design with one of the instructors before you start building.
- For maximum flexibility in positioning, removing, and reinstalling optics, use 3 cage rods instead of 4. Use a cage plate (CP02) for each lens. Mount the lens in a 1/2" lens tube (SM1L05) so it can be easily removed or installed.
- In the second half of the semester, you will upgrade your microscope to make two-color images. To avoid having to disassemble your microscope later, you'll leave an extra cage cube (C6W) in between your green LED and the excitation tube lens. Mounting the cube to a post will also provide some stability and rigidity to your illuminator.
- As before, the B1C cage cube cover plate has a shallow 8-32 threaded hole. Be sure to use 2 washers in between your TR2 post and the B1C base to be able to tighten the screw adequately without it getting stuck.
- Use the tiny 1/4" cage rods to connect the two C6W cages together
- Use a cube optic mount (B5C) to mount dichroic mirror DM on a kinematic platform (B4C). Place the kinematic platform in the cage cube (C6W).
- The first surface of the dichroic should face the laser.
- Some dichroics have an arrow indicating the first surface. If not, to ascertain which surface has the coating, watch the reflection of a corner of a piece of lens paper as it touches the mirror. On the first surface, the corner and its reflection will appear to touch. Held the other way, the corner and its image will appear a few millimeters apart.
- Mount the dichroic so that the first surface lies on exactly a diameter of the B5C mount.
- The mounting bracket should not stand in the way of the rotation of the kinematic stage.
- Use the clear plastic #4/40 screws affix the dichroic mirror without deforming nor scratching its surface.
- Hold the screwdriver at the tip to avoid slipping and scratching the dichroic. The screwdriver will instantly scratch the dichroic.
- Ask an instructor for help if you need to clean a dichroic mirror or barrier filter. Dichroics have delicate, exposed coatings and must be cleaned with extra care.
- The first surface of the dichroic should face the laser.
- Use four black, plastic #4/40 screws to hold the cube optic mount (B5C) on the cage cube (C6W).
- Tighten the screws enough so that the cube optic mount holds its adjustment, but can still be rotated.
- Make sure to block the (small percentage of) excitation light that will be transmitted through the dichroic mirror with some black lab tape affixed to the B5C cube optic mount.
Align the illumination path
- Use the lowest possible LED current when aligning to reduce risk of light getting in your eyes. Make sure to use some plastic pipe as a sleeve around the LED to reduce stray light.
- Remove jewelry. Turn of the LED before using reflective tools.
- Turn on the LED.
- Adjust the dichroic mirror DM so that the beam will enter the middle of the objective lens.
- Remove your objective and replace it with a DG10 mounted frosted-glass alignment disk with a pinhole to show the optical center of the vertical beam path.
- The B4C platform provides adjustment for the light path coming from your LED. Rotate it coarsely, or use the adjustment screws to center the beam in X and Y.
- Verify that the excitation light is centered on the objective and directed perpendicular to the floor. If the beam is at an angle, verify that the dichroic mirror DM is mounted on a diameter of the rotating mount.
- Roughly align the LED condenser lens to make an image near the back focal plane of the objective
- We don't know exactly where the BFP is, but it is roughly near the blue line on the 40x objective. We'll do some fine tuning to find the exact location later.
- Remove the DG10 frosted glass alignment tool. Use a piece of lens paper or a kimwipe and hold it roughly where you anticipate the BFP to be. Slide the green LED along its cage rods until an image forms on your paper. Make sure your excitation beam is still centered and vertical before proceeding.
- Use a fluorescence reference slide to center the field of view and to optimize the uniformity of illumination.
- Put a barrier filter in the afocal part of the imaging path.
- Center the camera's FOV in the objective's FOV.
- The barrier filter has delicate, exposed metal coatings. Ask an instructor for help if you need to clean it.
- The product of beam divergence and diameter is constant. L5 and L1 shrink the beam, causing increased divergence. The beam emerging from the objective will likely appear to grow in size as it propagates, even when the lenses are in their optimal positions.
Fluorescence imaging
In this part of the lab, you will make images of fluorescent microspheres and correct them for nonuniform illumination. In order to do the correction, you will need a reference image and dark image in addition to the image of the sample.
- Read this page to learn about the flat-field correction.
- Reference Images: Take the reference image as close as possible in time to the sample images. Don't make any adjustments to your microscope between capturing the reference image and the sample image. For example, every time you move a mirror or re-align the laser, you will change the illumination profile and you must take a new reference. Adjusting the camera exposure and gain between recording the reference and sample images is okay.
- Dark images: Each time you record an image (reference or sample), make sure to take a corresponding dark image using the EXACT SAME camera settings (i.e. use the exact same Exposure Time and Gain settings you had chosen for your reference/sample image). This is the only valid way to subtract the correct dark value from your reference/sample image. Use 12-bit imaging mode to get the best results.
- Read through this page to refresh you memory on saving in MATLAB. We recommend using the save command to save data as a .mat file that you can load back into MATLAB later. This makes it easy to reanalyze your images if necessary. You could also save in an image file format with imwrite. Convert the image to 16-bit, unsigned integer format with the correct range before saving.
Take images
- Record a reference image
- Put in an ND filter and take an image of the reference slide with the 40X objective.
- Turn off the laser and record a dark image (without changing any camera settings!)
- Use MATLAB to generate a histogram of to be certain that the image is exposed correctly.
- Remove the ND filter and image the samples:
- 0.84 μm beads and a corresponding dark image
- 3.26 μm beads and a corresponding dark image
- For each of these samples remember to make sure the image is exposed correctly, and to take a corresponding dark image
- Adjust the gain (to zero) and exposure of the camera to get the best picture.
Flat field correction
- Perform flat-field correction on the images.
- Divide the image by a normalized version of your reference image minus the dark image (see this page for more detail).
- Tip: If all your images are appearing dark in MATLAB you probably have not rescaled your image properly. Read through the displaying images section of this page if you need a refresher.