Line 280: Line 280:
 
According to theory,<ref>A. Einstein, [http://www.math.princeton.edu/~mcmillen/molbio/papers/Einstein_diffusion1905.pdf On the Motion of Small Particles Suspended in Liquids at Rest Required by the Molecular-Kinetic Theory of Heat], Annalen der Physik (1905).</ref><ref>E. Frey and K. Kroy, [http://www3.interscience.wiley.com/cgi-bin/abstract/109884431/ Brownian motion: a paradigm of soft matter and biological physics], Ann. Phys. (2005). Published on the 100th anniversary of Einstein’s paper, this reference chronicles the history of Brownian motion from 1905 to the present.</ref><ref>R. Newburgh, [http://scitation.aip.org/journals/doc/AJPIAS-ft/vol_74/iss_6/478_1.html Einstein, Perrin, and the reality of atoms: 1905 revisited], Am. J. Phys. (2006). A modern replication of Perrin's experiment. Has a good, concise appendix with both the Einstein and Langevin derivations.</ref><ref>M. Haw, [http://stacks.iop.org/JPhysCM/14/7769 Colloidal suspensions, Brownian motion, molecular reality: a short history], J. Phys. Condens. Matter (2002).</ref> the mean squared displacement of a suspended particle is proportional to the time interval as: <math>\left \langle {\left | \vec r(t+\tau)-\vec r(t) \right \vert}^2 \right \rangle=2Dd\tau</math>, where <i>r</i>(<i>t</i>) = position, <i>d</i> = number of dimensions, <i>D</i> = diffusion coefficient, and <math>\tau</math>= time interval.  
 
According to theory,<ref>A. Einstein, [http://www.math.princeton.edu/~mcmillen/molbio/papers/Einstein_diffusion1905.pdf On the Motion of Small Particles Suspended in Liquids at Rest Required by the Molecular-Kinetic Theory of Heat], Annalen der Physik (1905).</ref><ref>E. Frey and K. Kroy, [http://www3.interscience.wiley.com/cgi-bin/abstract/109884431/ Brownian motion: a paradigm of soft matter and biological physics], Ann. Phys. (2005). Published on the 100th anniversary of Einstein’s paper, this reference chronicles the history of Brownian motion from 1905 to the present.</ref><ref>R. Newburgh, [http://scitation.aip.org/journals/doc/AJPIAS-ft/vol_74/iss_6/478_1.html Einstein, Perrin, and the reality of atoms: 1905 revisited], Am. J. Phys. (2006). A modern replication of Perrin's experiment. Has a good, concise appendix with both the Einstein and Langevin derivations.</ref><ref>M. Haw, [http://stacks.iop.org/JPhysCM/14/7769 Colloidal suspensions, Brownian motion, molecular reality: a short history], J. Phys. Condens. Matter (2002).</ref> the mean squared displacement of a suspended particle is proportional to the time interval as: <math>\left \langle {\left | \vec r(t+\tau)-\vec r(t) \right \vert}^2 \right \rangle=2Dd\tau</math>, where <i>r</i>(<i>t</i>) = position, <i>d</i> = number of dimensions, <i>D</i> = diffusion coefficient, and <math>\tau</math>= time interval.  
  
*Track some 1&mu; and 2&mu; microspheres and estimate the diffusion coefficient.
+
*Track some 3&mu;m (Samples A & B) and 5&mu;m (Samples C & D) microspheres.
 +
*Estimate the diffusion coefficient of these microspheres suspended in solutions of varying viscosities. For reference, Sample A contains 3&mu;m spheres suspended in water.  
 
*Consider how many particles you should track and for how long. What is the uncertainty in your estimate?
 
*Consider how many particles you should track and for how long. What is the uncertainty in your estimate?
  

Revision as of 03:27, 23 September 2009

20.309: Biological Instrumentation and Measurement

ImageBar 774.jpg

1665 2009
Hooke-Microscope.png

Robert Hooke's microscope

20.309-Microscope.jpg

20.309 student's microscope

Hooke-CorkMicrograph.png

Hooke micrograph of cork cells

20.309-OnionMembranes.jpg

CCD image of fluorescent labeled intracellular membranes[1]

I took a good clear piece of Cork, and with a Pen-knife sharpen'd as keen as a Razor, I cut a piece of it off, and thereby left the surface of it exceeding smooth, then examining it very diligently with a Microscope, me thought I could perceive it to appear a little porous; but I could not so plainly distinguish them, as to be sure that they were pores, much less what Figure they were of: But judging from the lightness and yielding quality of the Cork, that certainly the texture could not be so curious, but that possibly, if I could use some further diligence, I might find it to be discernable with a Microscope, I with the same sharp Penknife, cut off from the former smooth surface an exceeding thin piece of it, and placing it on a black object Plate, because it was it self a white body, and casting the light on it with a deep plano-convex Glass, I could exceeding plainly perceive it to be all perforated and porous, much like a Honey-comb, but that the pores of it were not regular; yet it was not unlike a Honey-comb in these particulars.

Robert Hooke[2]


Don't you jusy buy a [expletive deleted] microscope?[3]

Anonymous 20.309 student[4]

Introduction

It would be difficult to overstate the role of optical microscopy in biological research. Since Robert Hooke noticed that the pores in a very thin sample of cork reminded him of the small rooms called cells where monks sleep in a monastery, the optical microscope has been a tool of central importance. But Hooke did more than notice the sample's texture. He also made perhaps the first quantitative measurements of cells with an optical microscope:

I told several lines of these pores, and found that there were usually about threescore of these small Cells placed end-ways in the eighteenth part of an Inch in length, whence I concluded there must be neer eleven hundred of them, or somewhat more then a thousand in the length of an Inch, and therefore in a square Inch above a Million, or 1166400. and in a Cubick Inch, above twelve hundred Millions, or 1259712000. a thing almost incredible, did not our Microscope assure us of it by ocular demonstration.

— Robert Hooke[2]

Improvements in optical components, microscope designs, illuminators, imaging devices, and sample preparation methodologies fostered increasingly sophisticated discoveries in the three and a half centuries since Micrographia. Barbara McClintock observed genetic transposition through an optical microscope by in 1944, for example. She was a talented microscopist who developed a technique that let her visualize and differentiate individual chromosomes in Zea Mays (corn) plant cells. McClintock was awarded the Nobel Prize in Physiology or Medicine in 1983 for this discovery. It took several decades before molecular techniques sufficiently sophisticated to confirm her discovery were developed.[5]

In this lab, you will design and build an optical microscope from components. You will characterize the performance of your instrument and make images using two different contrast methodologies: transmitted brightfield and epifluorescence. Using particle tracking computer software, you will make quantitative measurements of small suspended particles in Brownian motion. Finally, you will use your microscope to make quantitative measurements of vesicle trafficking in live plant cells. (In particular, you will estimate the number of ATP molecules per second required for intracellular transport of vesicles.)

In past semesters, several student groups have undertaken final projects that involved expanding the capabilities of their microscope, including implementing scanning-sample confocal and darkfield contrast modes.

Objectives

  • Learn about the theory and practice of light microscopy
  • Use ray tracing rules to design a transmitted bright field and fluorescent light microscope
  • Construct the microscope from optical components
  • Characterize the microscope's performance
  • Record, enhance, and analyze microscope images
  • Track moving particles
  • Make quantitative measurements of biological systems

How to do this lab

Week 1: microscope design and construction

  • Read the references; understand lenses, ray tracing, and magnification
  • Design and build microscope
  • Characterize the transmitted bright field performance of the microscope
  • Add a laser illumination beam path

Week 2: instrument characterization

  • Characterize the fluorescent imaging performance of the microscope
  • Image fluorescent samples
  • Correct images for nonuniform illumination

Week 3

  • Track microspheres suspended in a solvent
  • Estimate diffusion coefficients; compute Avagadro's number
  • Track the motion of vesicles under transport in a plant cell
  • Estimate the number of myosin motors involved in vesicle transport

Week 4: data analysis, image processing and report writing

Microscopy lab etiquette

During the microscopy lab, approximately two thousand optical components will be taken from stock, assembled into microscopes, and properly returnd to their assigned places. With the goal of keeping the frustration level to a minimum, please observe the following:

  • Observe laser safety guidelines.
  • Wear gloves when you are handling biological samples.
  • Store your microscope in one of the cubby holes in 16-317. If you use one of the high shelves, get somebody to help you lift.
  • Keep all of the boxes for the optics you use with your instrument to simplify putting things away.
  • Take a blue bin to store loose items (such as lens boxes) in.
  • Stages, CCD cameras, and barrier filters stay at the lab station. Do not store these with your microscope.
  • Do not move CCD cameras from one station to another.
  • Return objective lenses to the drawer when you are not using them. (Do not store them with your microscope.)
  • The stages are very expensive. Always lift from the bottom.
  • Ask a TA or instructor for help if you need to clean a dichroic or barrier filter. (These optics have exposed coatings and must be cleaned with extra care.)
  • Check the focal length of lenses before you use them. Some students before you have been very bad at the three-of-these-things game. If you find an optic in the wrong box: identify the optic and replace it in the correct box. (Ask an instructor or TA if you can't find the right box.)
  • Never use an SM1T2 coupler without a locking ring — they are very difficult to remove if they are tightened against a lens tube or tube ring.
  • Use tube rings (never an SM1T2, SM1V01, or SM1V05) to mount optics in lens tubes.
  • If you break something (or discover something pre-broken for you), do not return it to the component stock. Give all broken items to an instructor or TA.
  • Please do not remove parts from the TA microscope
  • Each group should be able to receive their own LED and dichroic mirror. Please ask a TA if you need one.
  • Many students have had trouble with where to begin as they design the microscope. Read through this lab manual carefully, many of the design constraints and tradeoffs have been outlined.
  • A good plan will often save you time and headaches down the road.

Readings and resources

Strongly suggested

Optical microscopy

Brownian motion

R. Newburgh, Einstein, Perrin, and the reality of atoms: 1905 revisited, Am. J. Phys. (2006). A modern replication of Perrin's experiment. Has a good, concise appendix with both the Einstein and Langevin derivations.

Other good sources

Brownian motion

A. Einstein, On the Motion of Small Particles Suspended in Liquids at Rest Required by the Molecular-Kinetic Theory of Heat, Annalen der Physik (1905).

M. Haw, Colloidal suspensions, Brownian motion, molecular reality: a short history, J. Phys. Condens. Matter (2002).

E. Frey and K. Kroy, Brownian motion: a paradigm of soft matter and biological physics, Ann. Phys. (2005).

Myosin XI and vesicle trafficking in plant cells

Tominaga, et. al. Higher plant myosin XI moves processively on actin with 35 nm steps at high velocity. The EMBO Journal Vol. 22 No. 6 pp. 1263-1272, 2003

S. Bolte, C. Talbot, Y. Boutte, O. Catrice, N. D. Read, B. Satiat-Jeunemaitre. FM-dyes as experimental probes for dissecting vesicle trafficking in living plant cells. Journal of Microscopy 214(2): 159-173 (May 2004).

Avisar, et. al. Myosin XI-K Is Required for Rapid Trafficking of Golgi Stacks, Peroxisomes, and Mitochondria in Leaf Cells of Nicotiana benthamiana1. Plant Physiology 146:1098-1108 (2008)

Peremyslov, V., et. al. Two Class XI Myosins Function in Organelle Trafficking and Root Hair Development in Arabidopsis. Plant Physiology 146:1109-1116 (2008)

Microscope design

Your microscope will be capable of two types of illumination: transmitted bright field and epi-fluorescence. It is probably easiest to design the trans illumination microscope first and then add a fluorescence capability.

Magnification

One of the simplest microscope consists of two lenses arranged in a 4-f configuration (as shown in the figure).

A 4-f lens system using lenses with focal lengths f1 and f2. The object and image distances (so and si, respectively) for the lenses are indicated.


Bright field transmitted microscopy is the simplest and most common optical microscopy method. In this technique, photons from an illuminator pass through the sample, where the may be absorbed, diffracted, or refracted. (The sample us usually mounted on a glass slide.) An objective lens on the opposite side of the sample collects the light. Most modern objective lenses produce collimated light, which is focused by a tube lens to form an image.

Illumination for epifluorescence microscopy reaches the sample through the objective lens — from the same side of the sample that is observed. Epi-fluorescence microscopy is normally used on samples that have been labled with a fluorescent molecule called a fluorophore. The (narrowband) illumination wavelength must match the absorption characteristic of the fluorophore. After becoming excited by a photon from the illuminator, fluorophores emit photons with a longer wavelength. A dichroic mirror in the microscope reflects the illumination wavelength but allows the emitted photons to pass through.


An example microscope made by the instructors will be available in the lab for you to examine. Feel free to make improvements on this design. Stability will be crucial for the particle tracking experiments. The required specification will be achieved through good design and careful construction — not by mindlessly overtightening screws.

Some elements must be positioned precise distances apart; other distances are not critical. Use ray-tracing to determine when this is the case.

Bright field transmitted illumination imaging

Sketch out a rough design for your microscope on paper. Begin with the bright field illumination path.

Note:

  • The Nikon objective lenses are designed to be paired with a 200 mm tube lens.
  • Assume that the objectives behave as ideal plano-convex lenses.
  • Fine focusing will be achieved by adjusting the height of the sample stage.
  • Start the alignment with a 10× objective but progress to 40× and 100×.
  • Use the LED illuminators for bright field transmitted light imaging.
  • Put a quick connect in your design such that the camera CCD will end up 200mm from the back focal plane of the objective. Remember that the CCD is recessed inside the opening of the camers.
  • The barrier filter has been installed in a lens tube already attached to the camera.

Fluorescence imaging

Now sketch a layout that includes both the bright field and fluorescence illumination paths.

  • Use a dichroic to direct the laser illumination toward the sample and to pass the emitted (fluorescent) light back through to the CCD.
  • The barrier filter removes any light from the illumination laser that was not reflected by the dichroic.
  • During fluorescent imaging, you will not use the bright field illuminator. Bright field capability is useful for first visualizing the sample and viewing what features are in the field of view. Your design should retain the ability to do both white light and fluorescent visualization.
  • Verify your design with one of the lab instructors before beginning construction.

Microscope components

Microscope Block Diagram

Rigid optical construction

The structure of your microscope will be built from a combination of cage and lens tube components from ThorLabs. (See the ThorLabs online catalog for more details. Print catalogs are available in the lab.) Be sure you understand how to use cage cubes (C4W), cube optic mounts (B5C), and kinematic mounting plates (B4C). Please ask about any components you are not sure how to use.

An example microscope put together by the instructors will be available in the lab for you to look at. Feel free to make improvements on this design. Stability will be crucial for the particle tracking experiments. This will be achieved through good design and careful construction — not by mindlessly overtightening screws.

Simple lenses

Plano-convex spherical lenses are available with focal lengths of 25, 50, 75, 100, 125, 150, 175, and 200 mm. It is best to mount most optics in short (0.5" or 0.3") lens tubes. It is acceptable to mount a lens between the end of a tube and a tube ring or between two tube rings. In most cases, the convex side of the lens faces toward the collimated beam; the planar side goes toward the convergent rays. Plano-concave lenses with focal lengths of -30 and -50 are also available.

  • Advice: Some students in the past have had difficulty with the Three of These Things game. Verify all optics before you use them by measuring the focal length with a ruler.

As you install lenses into your microscope, put a piece of tape on the lens tube showing focal length and orientation. This will help you both during constructino and put-away. Save the lens storage boxes and return components to the correct boxes when you are done.

Handle lenses only by the edges. If a lens is dirty, first remove grit with a blast of clean air or CO2. Clean the lens by wiping with a folded piece of lens paper wetted with a drop of methanol. (Do not touch the part of the tissue you use for cleaning with your fingers.) In some cases, it may be helpful to hold the folded lens tissue in a hemostat. Ask an instructor if you need help.

Objective lenses

Please see the Nikon Introduction to Microscope Objectives at their excellent MicroscopyU website.

There are three objective lenses available in the lab: a 10×, a 40×, and a 100×. All of these are designed for a 200 mm tube lens. An adapter ring converts the objective mounting threads to the SM1 threads used by the lens tube system.

  • The back focal plane (BFP) of the objective coincides with the rear of the objective housing. This is equivalent to the focal plane of a simple lens.
  • Working distance (WD) is the distance between the front end of the objective and the sample plane (when

the sample is in focus). Generally, the higher the magnification, the lower the working distance.

  • The 100× objective is designed to be used with immersion oil, which provides an optical medium of pre-

determined refractive index (n = 1.5). When using the 100× objective, place a drop of oil on it. Bring the drop in con- tact with the slide cover glass. After use, clean off excess oil by wicking it away with lens paper. Do not put samples away dirty. It is not necessary to use immersion oil for thin samples such as the Air Force Target or Ronchi Ruling.

Sample stage

Precision Newport X/Y/Z stages mounted on a post are available at each lab station. The stage setup is top-heavy. Avoid accidents by ensuring that the post base is always attached to an optical breadboard or table. Return the stage to the lab station when you are done with it.

The z-axis adjustment of the sample stage provides fine focusing.

Fluorescence illumination

The fluorescent illumination source is a 5 mW, λ=532 nm green laser pointer. Do not begin working with the laser until you are familiar with laser safety procedures. If you missed laser safety training, please see an instructor.

  • Wear the correct laser safety eyewear at all times — even when your laser is not on — since other groups may be using theirs.
  • Make sure the laser warning light outside the main door to the lab is flashing before you turn on a laser.
  • Never point the laser toward other people.
  • Laser light can emerge from the top of the objective lens. Never put your face directly above the objective.

Beam expansion

Collimated light comes out of the laser pointer. The light should also be collimated when it reaches the sample (Kohler illumination). In order to provide a good image, the laser light should illuminate most of the field of view. The beam emerging from the laser pointer (abuot 1.1 mm) may not be wide enough. This means that the laser beam will need to be expanded. What expansion factor should you use for a 40× objective? The design of the beam expander is a tradeoff. Overexpansion will decrease the light intensity at the sample and may not give you enough power for imaging. Underexpansion will result in very uneven illumination.

Dichroic mirror and barrier filter

The fluorophores we will emit light in the orange-red region of the visible spectrum (550-600 nm) when excited by the 532 nm green laser.

In any fluorescence system, a key concern is viewing only the emitted fluorescence photons, and eliminating any background light, especially from the illumination source. Two optical elements address the problem. A dichroic mirror passes light of one wavelength, and reflects light of another. The transmission spectrum for the 565DCXT dichroicis shown in Fig. 3. A barrier filter blocks a particular spectral region. We will use the E590LPv2 barrier filter.

Figure 1 shows the tranmittance of the dichroic mirror and barrier filter over a range of wavelengths. The barrier filter is essential for high-sensitivity fluorescence imaging. It will pass the red light from the LED in the illuminator for bright field imaging as well as the fluorescence photons.

Figure 3: The transmission spectra for the 565DCXT dichroic and the E590LPv2 barrier filter

CCD camera

The microscope you will build does not have an eyepiece for direct visual observation. Instead, images will be captured with a Firewire-enabled CCD camera (DMK 21F04 from The Imaging Source). Its monochrome (black and white) sensor contains a grid of 640×480 square pixels that measure 5.6 μm on a side. An adapter ring converts the C-mount thread on the camera to SM1.

The barrier filter, SM1 adapter, and a quick-connect have been installed on the camera. Please leave the camera at the lab station when you are done with it. Camera software called IC Capture is installed on the labstation PCs.


Experiment 1:microscope construction and characterization

The microscope should be built in two stages. First, build a white-light inverted microscope. Verify its alignment and magnification.

Next, add the fluorescence branch. An adjustable iris aperture and piece of lens or tissue paper can be very helpful for aligning the laser and directing it along the axis of the tubes.

Bright field calibration

Make images the following samples using the three different objectives (10×, 40×, and 100×):

  • The smallest line pair on the Air Force imaging target
  • A slide of 3.2μm or 1μm latex spheres
  • Ronchi ruling - a periodic pattern containing 600 line-pairs per mm

Can you see all these samples? What is the magnification of the microscope and the size of its field of view? Is it what you expected?

Experiment 2:Fluorescence characterization and imaging

You have the following samples available for imaging using both 40× and 100× objectives:

  • A fluorescence reference slide
  • A sample slide with 4 μm red-fluorescent beads (modified with Nile Red dye, peak excitation at 535nm, peak emission at 575nm).

Imaging tasks:

  1. Use a fluorescence reference slide to center the field of view and to optimize the uniformity of illumination. Take an image of this. Use the histogram display of the camera software to be certain that the image is not overexposed.
  2. Take an image of the fluorescent beads. Perform flat field correction on the image(i.e. divide the image by the normalized reference image). Compare what you see before and after flat field correction.
  3. Image a stained onion slide in bright field and fluorescent contrast. Do a flat-field correction on the fluorescent image. Overlay the fluorescent image on top of the bright field image.
  4. Make an image of the PSF slide. This slide consists of 190nm fluorescent beads plus some 1μ beads to help focus in bright field contrast. You may want to make multiple images of the slide and average them to reduce noise. If you do so, be sure not to bump the table or make any adjustments in-between images.

FM-dye reference[6]

Experiment 3: tracking particles in suspension

Introduction and background[7]

If you have ever looked at an aqueous sample through a microscope, you have probably noticed that every small particle you see wiggles about continuously. Robert Brown, a British botanist, was not the first person to observe these motions, but perhaps the first person to recognize the significance of this observation. Experiments quickly established the basic features of these movements. Among other things, the magnitude of the fluctuations depended on the size of the particle, and there was no difference between "live" objects, such as plant pollen, and things such as rock dust. Apparently, finely crushed pieces of an Egyptian mummy also displayed these fluctuations.

Brown noted: [The movements] arose neither from currents in the fluid, nor from its gradual evaporation, but belonged to the particle itself.

This effect may have remained a curiosity had it not been for A. Einstein and M. Smoluchowski. They realized that these particle movements made perfect sense in the context of the then developing kinetic theory of fluids. If matter is composed of atoms that collide frequently with other atoms, they reasoned, then even relatively large objects such as pollen grains would exhibit random movements. This last sentence contains the ingredients for several Nobel prizes!

Indeed, Einstein's interpretation of Brownian motion as the outcome of continuous bombardment by atoms immediately suggested a direct test of the atomic theory of matter. J. Perrin received the 1926 Nobel Prize for validating Einstein's predictions, thus confirming the atomic theory of matter.

Since then, the field has exploded, and a thorough understanding of Brownian motion is essential for everything from polymer physics to biophysics, aerodynamics, and statistical mechanics. One of the aims of this lab is to directly reproduce the experiments of J. Perrin that lead to his Nobel Prize. A translation of the key work is included in the reprints folder. Have a look – he used latex spheres, and we will use polystyrene spheres, but otherwise the experiments will be identical. In addition to reproducing Perrin's results, you will probe further by looking at the effect of varying solvent molecule size.

Microscope stability for particle tracking

To verify that your system is sufficiently stable for accurate particle tracking, measure a dry specimen containing 1μ beads in bright field contrast. Chose a field of view in which you can see at least 3-4 beads. Using a 40x objective, track the beads for about 3 min. Use the bead tracking processing algorithm on two beads to calculate two trajectories. To further reduce common-mode motion from vibrations, calculate the differential trajectory from the individual trajectories of these two beads. Calculate the MSD from the differential trajectory. Your MSD should start out less than 10 nm2 at t = 1 sec. and still be less than 100 nm2 for t = 180 sec.

Estimating the diffusion coefficient by tracking suspended microspheres

According to theory,[8][9][10][11] the mean squared displacement of a suspended particle is proportional to the time interval as: $ \left \langle {\left | \vec r(t+\tau)-\vec r(t) \right \vert}^2 \right \rangle=2Dd\tau $, where r(t) = position, d = number of dimensions, D = diffusion coefficient, and $ \tau $= time interval.

  • Track some 3μm (Samples A & B) and 5μm (Samples C & D) microspheres.
  • Estimate the diffusion coefficient of these microspheres suspended in solutions of varying viscosities. For reference, Sample A contains 3μm spheres suspended in water.
  • Consider how many particles you should track and for how long. What is the uncertainty in your estimate?

See: this page for more discussion of Brownian motion and a Matlab simulation.

Experiment 4: Transport in a plant cell[7]

The eukaryotic cytoskeleton. Actin filaments are shown in red, microtubules in green, and the nuclei are in blue.

In the fourth part of this experiment, you will characterize the motion of particles inside a living cell.

An out-of-date view of a cell was that it is a "bag of water" containing various enzymes in which matter is transported passively by diffusion. Though diffusion is an important mechanism, it is too slow and random for long distance transport and directing materials where they are most needed, especially in larger cells. It is now understood that cells have highly developed and intricate mechanisms for directed transport of materials.

Cells transport food, waste, information, etc. in membrane-bound vesicles, which are visible under a light microscope.

Most motions within and of cells involve two components, a cytoskeletal fiber that serves as a track, and a motor protein that does the work. The motor molecule uses energy from the hydrolysis of one ATP molecule to bind to the fiber, bend to pull itself along the fiber, and release, all of which constitutes one "step". For an animation of this stepping process, see this movie animation from the Vale lab web site at UC San Francisco. One can divide cellular motility mechanisms into two classes based on the cytoskeletal fibers involved. Microtubule-based mechanisms involve dynein or kinesin motors pulling on microtubules made of the protein tubulin. Actin-based mechanisms involve myosin motors pulling on actin fibers, also called microfibers.

Cartoon of myosin motors pulling organelles along an actin filament.
Binding of kinesin motor to microtubule.

Virtually all cell types exhibit directed intracellular transport, but some cell types are particularly suitable for transport studies. Fish-scale pigment cells work well, since a large fraction of the cargoes that are transported are pigmented and thus easy to observe – the disadvantage is that you would need to bring a living fish into lab as a source of these cells. For convenience, we will use epidermal cells from onion bulbs that you can easily acquire in a grocery store. With some care, a single layer of cells can be peeled off an inner section of the onion bulb and mounted flat on a slide.

Onion cells in bright-field illumination. Round object in each cell is the nucleus.
Vesicles in the cytoplasm of a plant cell, as seen in dark-field.

In this experiment, we will be viewing the movement of vesicles within the cytoplasm of onion epidermal cells, shown above as they appear in bright-field and dark-field microscopy. The layers you see in an onion bulb develop into leaves when it sprouts. Both sides of the leaf are covered with an epidermis consisting of brick-shaped cells, each with a cell wall and cell membrane on the outside.

Most of the interior of the cell is filled with a clear vacuole that functions in storage and in maintenance of hydrostatic pressure essential to the stiffness of the plant (the difference between crisp lettuce and wilted lettuce). The cytoplasm, containing all of the other cell contents, occurs in a thin layer between the cell membrane and the vacuole, and in thin extensions through the vacuole called transvacuolar strands. It is within the cytoplasm that you will be observing directed transport of vesicles by an actin-based mechanism.

These vesicles are spherical or rod-shaped organelles such as mitochondria, spherosomes, and peroxisomes ranging in size from 0.5 to 3 microns. The diagram of an onion cell below shows the location of the cell wall, cytoplasm and vesicles in a typical cell; you will not be able to see much of the endoplasmic reticulum or the vacuole depicted because of their transparency. Under the microscope, you will notice the vesicles are located just along the edges of the cell, or near the top and bottom surface if you focus up and down. When you see a narrow band of moving vesicles in the center of the cell, it is located in a transvacuolar strand, which may be a handy place to study motion.

A 3D cross-section model of an onion epidermal cell, showing actin filaments and vesicles in the narrow bands of cytoplasm within the cell.

In plant cells, vesicles are transported along actin fibers by myosin motor molecules. An actin filament is composed of two intertwined actin chains, about 7 nm in diameter. An actin fiber is structurally polar, having a (+) end and a (-) end. Most myosin motors move towards the (+) end of the actin fiber. In order to reverse the direction of a vesicle's motion, the vesicle must grab on to another actin fiber oriented in the opposite direction.

There are at least eighteen described classes of myosin, of which three, myosin VIII, XI, and XII are found in plant cells. Some myosin motors are processive, meaning that they remain bound to an actin fiber as they move step-by-step along it (analagous to this movie animation of kinesin). Other myosins are non-processive, releasing from the actin fiber after each step. Myosin II found in muscle cells is non-processive, as illustrated in this video animation. In the muscle functional unit, there are many myosin motors acting together, so there are always some attached to the actin fiber. The myosin XI responsible for transport of plant cell vesicles is the fastest myosin known and is processive. It apparently is not known how many myosin molecules are attached to the surface of a vesicle or how many of those are active at one time in pulling the vesicle along an actin fiber.

In some plant cells and algal cells, a large-scale streaming motion of the cytoplasm is observed, logically called cytoplasmic streaming. This bulk flow is believed to be caused by myosin motors pulling the extensive endoplasmic reticulum along actin fibers lining the cell membrane. Many other vesicles are then dragged along with the endoplasmic reticulum. Lodish and Berk, et al. provide a detailed explanation of this process and a video of cytoplasmic streaming in the pond eeed Elodea can be viewed here.

In your observations of vesicles in onion epidermal cells, you should distinguish between the random Brownian motion of vesicles that are unattached (or at least not actively moving along) actin filaments, the directed transport of vesicles by attached myosin motors, and possibly (though we are not sure this really happens in onions) bulk flow of vesicles in cytoplasmic streaming.

Report Requirements

Microscope documentation

Make a sketch or block diagram of your full microscope setup (a hand drawing is perfectly acceptable, but please keep it neat - a ruler is handy) with important parameters indicated (i.e. lens focal lengths, distances of main components, etc.). It is not necessary to document the structural components.

Bright field

  1. Your favorite bright field images — one or two with each objective. Indicating the magnifcation and feld of view. Do they match what was expected?
  2. Calculate and include a table of the diffraction-limited resolution for the 10X, 40X and 100X objectives.

Fluorescence

  1. Filtered, normalized image you used for flat field correction with histogram and plot of one horizontal line near the middle.
  2. Uncorrected and flat-field corrected fluorescent image of 4μ beads.
  3. Image of PSF slide with 40X objective. Estimate the resolution using this objective, assuming the 190nm beads have negligible size. (Is this a good assumption? Can you improve on it?) One approach is to fit a Gaussian function to the image of one particle.
  4. Compare your estimate of the resolution of the 40X objective to the value you calculated.

Particle tracking

  1. Estimate the diffusion coefficient of the 1μ and 2μ microspheres using the particle tracks you recorded.
  2. Estimate the velocities of the granules being transported in the onion cell.


References

  1. Onion endothelial cell incubated with FM 4-64 dye (Invitrogen). See class stellar site for protocol. Oh & Yamaguchi, unpublished lab report
  2. 2.0 2.1 Hooke, R. Micrographia: or Some Physiological Descriptions of Minute Bodies made by Magnifying Glasses with Observations and Inquiries Thereupon London:Jo. Martyn, and Ja. Allestry, Printers to the Royal Society; 1665
  3. http://www.historycommons.org/context.jsp?item=a043074editedtranscripts
  4. Fall 2007
  5. See, for example: McClintock, B. The origin and behavior of mutable loci in maize. PNAS. 1950; 36:344-355. [1], [2], and Endersby, Jim. A Guinea Pig's History of Biology. Cambridge, Massachusetts: Harvard University Press; 2007.
  6. S. Bolte, C. Talbot, Y. Boutte, O. Catrice, N. D. Read, B. Satiat-Jeunemaitre. FM-dyes.pdf FM dyes as experimental probes for dissecting vesicle trafficking in living plant cells. Journal of Microscopy 214(2): 159-173 (May 2004).
  7. 7.0 7.1 taken from http://labs.physics.berkeley.edu/mediawiki/index.php/Brownian_Motion_in_Cells
  8. A. Einstein, On the Motion of Small Particles Suspended in Liquids at Rest Required by the Molecular-Kinetic Theory of Heat, Annalen der Physik (1905).
  9. E. Frey and K. Kroy, Brownian motion: a paradigm of soft matter and biological physics, Ann. Phys. (2005). Published on the 100th anniversary of Einstein’s paper, this reference chronicles the history of Brownian motion from 1905 to the present.
  10. R. Newburgh, Einstein, Perrin, and the reality of atoms: 1905 revisited, Am. J. Phys. (2006). A modern replication of Perrin's experiment. Has a good, concise appendix with both the Einstein and Langevin derivations.
  11. M. Haw, Colloidal suspensions, Brownian motion, molecular reality: a short history, J. Phys. Condens. Matter (2002).