Before you record any images, it is a very good idea to adjust your microscope optics to make the illumination is as good as possible. Even if you adjust the illuminator perfectly, the light intensity won't be exactly the same everywhere in the field of view. In most cases, the illumination is brightest in the middle of the image, with a steady decrease in intensity as you approach the edges of the field of view. Dirt or other flaws in the illumination system can also cause variation in the illumination. Another source of non-uniformity is pixel-to-pixel variation in the dark level and responsivity of the camera. Because of this, even in complete darkness, all of the pixels on the detector will not necessarily report the same value. Identical light intensities may produce different outputs depending on which pixel on the detector records the signal. These problems can be largely eliminated through a technique called flat-field correction or shading correction.

Correcting an image

The effect of nonuniform illumination and camera response can be reduced by flat-field correction, which involves dividing the non-uniformly illuminated image by a reference image that maps the illumination intensity. The black level must be subtracted from image to be corrected and the reference image in order to apply the proper correction.