Flat-field correction

From Course Wiki
Revision as of 01:16, 8 September 2015 by Steven Wasserman (Talk | contribs)

(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search
20.309: Biological Instrumentation and Measurement

ImageBar 774.jpg


Overview

Before you record any images from the camera, it is a very good idea to adjust your microscope optics to make the illumination is as good as possible. Even if you adjust the illuminator perfectly, the light intensity won't be exactly the same everywhere in the field of view. In most cases, the illumination is brightest in the middle of the image, with a steady decrease in intensity as you approach the edges of the field of view. Dirt or other flaws in the illumination system can also cause variation in the illumination. Another source of non-uniformity is pixel-to-pixel variation in the dark level and responsivity of the camera. This means that, even in complete darkness, the camera will report a range of values, and identical light intensities may produce different results depending on where they are recorded on the detector. These problems can be largely eliminated through a technique called flat-field correction or shading correction.

Correcting an image