The frequencies here are not properly calibrated until I get the DAQ bandwidth.  There are obvious peaks at what are (now) 10, 20, and 40 Hz -- I think these are the harmonics of the e coli rotation, and once the abscissa is properly scaled will give the e coli rotation speed.  There looks like there is also a peak hidden in the DC peak.
  
 
===DNA Tethers===
 
===DNA Tethers===
 
[[File:dna_raw.png]]
 
[[File:dna_raw.png]]

Revision as of 04:17, 2 March 2012

Calibration

All calibrations and experiments done at 20 mW.

Position Calibration

1um cal.png calibration: 502463 Volts / m = 1.99 um / volt

Trap Stiffness Calibration

Method Trap Stiffness (pN / nm)
Stokes 2.26e-5 pN/nm
Equipartition Theorem 1.23e-5 pN/nm
PSD *Need Bandwidth of DAQ*

Experiments

E Coli

Ecoli.png

The frequencies here are not properly calibrated until I get the DAQ bandwidth. There are obvious peaks at what are (now) 10, 20, and 40 Hz -- I think these are the harmonics of the e coli rotation, and once the abscissa is properly scaled will give the e coli rotation speed. There looks like there is also a peak hidden in the DC peak.

DNA Tethers

Dna raw.png