Difference between revisions of "Assignment 5, Part 1: viscosity and diffusivity in glycerol-water mixtures"
From Course Wiki
(→Estimating the diffusion coefficient by tracking suspended microspheres) |
(→Estimating the diffusion coefficient by tracking suspended microspheres) |
||
Line 18: | Line 18: | ||
:''Tip 2'': Each full frame of the full camera field of view takes up almost 2.5 MB of memory, so movies can get large very fast. Try to limit image data variables to a reasonable size by keeping the length of the movie short or limiting the Region of Interest (ROI) to a fraction of the full field of view.( A full-field, three minute move takes up about 4 GB, which is certain to push MATLAB over the edge.) | :''Tip 2'': Each full frame of the full camera field of view takes up almost 2.5 MB of memory, so movies can get large very fast. Try to limit image data variables to a reasonable size by keeping the length of the movie short or limiting the Region of Interest (ROI) to a fraction of the full field of view.( A full-field, three minute move takes up about 4 GB, which is certain to push MATLAB over the edge.) | ||
− | 2. Record movies of beads diffusing in the two glycerol solutions and use your newly developed code to estimate the diffusion coefficient of each sample. | + | 2. Record movies of beads diffusing in the two glycerol solutions and use your newly developed code to estimate the diffusion coefficient of each sample. '''Consider how many particles you should track and for how long. What factors determine the uncertainty and accuracy of your estimate?''' |
− | + | ||
− | + | ||
{{Template:Assignment Turn In|message=Turn in for the viscosity part: | {{Template:Assignment Turn In|message=Turn in for the viscosity part: | ||
Line 28: | Line 26: | ||
# Plot the average MSD vs τ results for the two glycerin samples (A and B); use log-log axes. Use the minimum number of axes that can convey your results clearly. | # Plot the average MSD vs τ results for the two glycerin samples (A and B); use log-log axes. Use the minimum number of axes that can convey your results clearly. | ||
# Include a table of the diffusion coefficient, viscosity and measured glycerin/water ratio for each of the samples (A and B) | # Include a table of the diffusion coefficient, viscosity and measured glycerin/water ratio for each of the samples (A and B) | ||
− | # Is the viscosity you measured close to the theoretical value predicted by | + | # Is the viscosity you measured close to the theoretical value predicted by [http://www.met.reading.ac.uk/~sws04cdw/viscosity_calc.html this website]? |
# Include a thorough discussion of error sources and your approaches to minimize them. It may be helpful to list out the error sources in a table, including a category for the error source, type of error (random, systematic, fundamental, technical, etc.), the magnitude of the error, and a description and way to minimize each one. | # Include a thorough discussion of error sources and your approaches to minimize them. It may be helpful to list out the error sources in a table, including a category for the error source, type of error (random, systematic, fundamental, technical, etc.), the magnitude of the error, and a description and way to minimize each one. | ||
}} | }} |
Revision as of 18:12, 6 March 2019
This is Part 1 of Assignment 5.
Estimating the diffusion coefficient by tracking suspended microspheres
1. Track some 0.86μm Pink Spherotech polystyrene beads in water-glycerin mixtures (Samples A and B contain 80% and 85% glycerin, respectively).
- Notes: Fluorescent microspheres have been mixed for you by the instructors into water-glycerin solutions A and B. (a) Vortex the stock Falcon tube, and then (b) transfer the bead suspension into its imaging chamber (consisting of a microscope slide, double-sided tape delimiting a 2-mm channel, and a 22mm x 40mm No. 1.5 coverslip, and sealed at both ends nail polish).
- Tip 1: Ensure that the focal plane you choose to image is not near the coverslip or the slide. If some particles don't move or hardly move at all, it is likely that they are stuck to the coverslip. Adjust the focus so that you are viewing a plane near the middle of the sample. (A good way to do this is to focus on the top and bottom of the sample chamber and then split the difference.)
- Tip 2: Each full frame of the full camera field of view takes up almost 2.5 MB of memory, so movies can get large very fast. Try to limit image data variables to a reasonable size by keeping the length of the movie short or limiting the Region of Interest (ROI) to a fraction of the full field of view.( A full-field, three minute move takes up about 4 GB, which is certain to push MATLAB over the edge.)
2. Record movies of beads diffusing in the two glycerol solutions and use your newly developed code to estimate the diffusion coefficient of each sample. Consider how many particles you should track and for how long. What factors determine the uncertainty and accuracy of your estimate?
Turn in for the viscosity part:
|